Propaq® CSVital Signs MonitorService Manual
6 Overview Welch Allyn Propaq CS Vital Signs MonitorWarranty serviceDO NOT ATTEMPT TO REPAIR the monitor yourself during the warranty period. For serv
96 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorInstalling the printer1. Lay the Chassis on its back2. Completely open the paper door.3.
Service Manual Repair procedures 97Opening the SpO2 Module1. Separate the SpO2 module from the monitor according to the procedure “Removing the dual b
98 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorReplacing the SpO2 BoardsThe following instructions describe how to replace the two SpO2
Service Manual Repair procedures 99Replacing the MSP/SpO2 BoardsThe following instructions describe how to replace the MSP/SpO2 Boards in the Expansio
100 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor• set the two boards and side panels in place, making sure that the boards are seated o
Service Manual Repair procedures 101Removing the SSP/Stick Board assembly1. Remove the two screws located on the right edge of the SSP Board (as viewe
102 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorRemoving the Stick Board from the SSP Board1. Disconnect the tubing from the pump and p
Service Manual Repair procedures 103Replacing Expansion Module side panelsThe side panels easily slide out after the Chassis and rear panel are separa
104 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor
7105Technical overviewIntroductionThis section describes the electronics of the Propaq CS monitor and three of its options: Printer, Pulse Oximetry (S
Service Manual Overview 7Identifying Propaq CS Monitor configurationsThe following tables identify Propaq CS monitor configurations and how they are i
106 Technical overview Welch Allyn Propaq CS Vital Signs MonitorMonitor options0696-12Rear ChassisRecharger BoardRight Side PanelLeft Side PanelAnalog
Service Manual Technical overview 1071097-04Expansion ModuleOption 221printerFiller PanelPrinterMain HousingPrinter DoorFiller PanelRear PanelFront Pa
108 Technical overview Welch Allyn Propaq CS Vital Signs Monitor1097-08Expansion ModuleOption 226SpO2SCO2 SpO2 Connector Panel Rear Panel Speake
Service Manual Technical overview 109hardware layout1097-10Expansion ModuleOption 228SpO2MCO2SCO2 SpO2 Connector Panel Rear Panel Speaker Panel08
110 Technical overview Welch Allyn Propaq CS Vital Signs Monitormonitor system system blockEXTENDERBOARDSPO2 BOARDSCP BOARDSPO2 ANALOG INPUTSPO2 ANALO
Service Manual Technical overview 111Power electronics descriptionThe power system circuits are located on the Analog, Digital, and Recharger Boards.
112 Technical overview Welch Allyn Propaq CS Vital Signs Monitorpower el ectronic s systemDISPLAYCONTROLLERBATTERYRECHARGERSUPPLYBATTERYMONITORPOWERCO
Service Manual Technical overview 113Monitoring electronics system descriptionThe monitoring electronics reside on the Analog and Digital Boards locat
114 Technical overview Welch Allyn Propaq CS Vital Signs MonitorDigital board interconnect diagram (2.4 GHz radio card)
Service Manual Technical overview 115Digital board interconnect diagram (5 GHz radio card)
8 Overview Welch Allyn Propaq CS Vital Signs MonitorTable 2. Expansion module option configurationsOption numberPrinter Mainstream CO2Sidestream CO2S
116 Technical overview Welch Allyn Propaq CS Vital Signs MonitorNIBP pneumatics descriptionThe NIBP (cuff) system:• inflates the cuff to a target cuff
Service Manual Technical overview 117CUFFPRIMARYTRANSDUCERSECONDARYTRANSDUCERDUMPVALVEOPENSERVO(LINEAR)VALVECLOSEDBLEEDVALVEOPENPUMPOFFATMOSPHERIC PRE
118 Technical overview Welch Allyn Propaq CS Vital Signs MonitorCUFFPRIMARYTRANSDUCERSECONDARYTRANSDUCERDUMPVALVECLOSEDSERVO(LINEAR)VALVECLOSEDBLEEDVA
Service Manual Technical overview 119CUFFPRIMARYTRANSDUCERSECONDARYTRANSDUCERDUMPVALVEOPENSERVO(LINEAR)VALVECLOSED*BLEEDVALVEOPENPUMPOFFATMOSPHERIC PR
120 Technical overview Welch Allyn Propaq CS Vital Signs MonitorCabling diagramssystem cabling diagram825-0146-00 AREF DWG, CABLING, 24X G SERIESMANAL
Service Manual Technical overview 121Expansion Module and printer descriptionThe printer shares battery power with the monitor, but contains its own p
122 Technical overview Welch Allyn Propaq CS Vital Signs MonitorPRINTERPOWERSYSTEMPOWERPRINTHEADPRINTERDRIVERPRINTERCPUPRINTERLOGICP150FROMBATTERYP1EX
Service Manual Technical overview 123Pulse oximetry option (SpO2) descriptionThe monitor pulse oximetry option (SpO2) hardware includes two circuit bo
124 Technical overview Welch Allyn Propaq CS Vital Signs Monitor• internal preamplifier and switch• ac-coupled amplifier• inverting amplifier with var
Service Manual Technical overview 125spo2 blockPULSE OXIMETRY BLOCK DIAGRAMINTERNALPREAMPAC COUPLED AMPPEAK DETECTORINVERTING AMP(VAR. GAIN STAGE)PEAK
Service Manual Overview 9Identifying the Acuity Communication optionsMonitor controlsSelf test and service menusThe Propaq CS monitor includes built-i
126 Technical overview Welch Allyn Propaq CS Vital Signs MonitorCapnography (CO2) descriptionTwo capnography (CO2) monitoring options are available—th
Service Manual Technical overview 127Mainstream BoardThe MSP Board is divided into isolated and non-isolated circuitry. Non-Isolated Circuitry. The No
128 Technical overview Welch Allyn Propaq CS Vital Signs MonitorIPC expansionThe Inter-Processor Communications (IPC) bus, originating in the monitor,
8129Field Replaceable Units (FRUs)To order a field-replaceable unit, contact Welch Allyn Technical Support as described in page ii.Replacement part de
130 Field Replaceable Units (FRUs) Welch Allyn Propaq CS Vital Signs MonitorNIBP HOSE FITTING, PLASTIC, PAIR 020-0065-00NIBP CONNECTOR, HOSE END, NICK
Service Manual Field Replaceable Units (FRUs) 131SCP BOARD, NELLCOR MP-203 ONLY, U11 64K PROM 020-0476-00SCP BOARD, NELLCOR / MASIMO, U303 128K PROM 0
132 Field Replaceable Units (FRUs) Welch Allyn Propaq CS Vital Signs MonitorLABEL, RIGHT SIDE PANEL, ACUITY (NO LONGER AVAILIBLE) 640-0326-00LABEL, RI
Service Manual Field Replaceable Units (FRUs) 133LABEL, PRINTER KEY PANEL 640-0465-00Special service toolsNIBP CALIBRATION SETUP TUBING ADAPTER KIT 00
134 Field Replaceable Units (FRUs) Welch Allyn Propaq CS Vital Signs Monitor
A135Manufacturable test equipmentThe following device(s) can be manufactured for use in the functional verification and calibration procedures in this
10 Overview Welch Allyn Propaq CS Vital Signs MonitorThe Propaq CS monitor contains software routines that make the Functional Verification and Calibr
136 Manufacturable test equipment Welch Allyn Propaq CS Vital Signs Monitor
B137Fluke patient simulator modificationThe factory configuration of the Fluke (Dynatech/Nevada) model 214A, 215A, and 217A Patient Simulators is inco
138 Fluke patient simulator modification Welch Allyn Propaq CS Vital Signs MonitorIncompatibilities with other simulators may also cause erroneous rea
Service Manual Overview 11Service test screensThere are two Acuity Wireless options available:1. 2.4 GHz radio card (002-0150-00)2. 5 GHz radio card (
12 Overview Welch Allyn Propaq CS Vital Signs MonitorNetwork Monitor screen (2.4 GHz radio card)To access this screen, from the Network Status screen
313Functional verificationIntroductionThis functional verification procedure helps to confirm the proper operation of the Propaq CS monitor and option
14 Functional verification Welch Allyn Propaq CS Vital Signs MonitorRequired equipmentYou will need the following equipment to verify that all functio
Service Manual Functional verification 15Safety testsThe following two safety tests, a risk (leakage) current safety check and a dielectric strength i
ii Welch Allyn Propaq CS Vital Signs MonitorCopyright 2008 Welch Allyn. All rights are reserved. No one is permitted to reproduce or duplicate, in any
16 Functional verification Welch Allyn Propaq CS Vital Signs MonitorDielectric strength (Hi-Pot) testBecause of the close spacings of the monitor&apos
Service Manual Functional verification 17Functional verificationPerform the functional verification with the monitor fully assembled. If the monitor h
18 Functional verification Welch Allyn Propaq CS Vital Signs Monitordc power acce ssory cable10. Turn on the dc power supply.11. Check that the green
Service Manual Functional verification 19System testsThe following procedures check that the buttons operate properly, that the display works correctl
20 Functional verification Welch Allyn Propaq CS Vital Signs Monitor15. Press ECG/RESP > MORE. Then press CHANGE to select the four HR/PR TONE leve
Service Manual Functional verification 21If incorrect, press NEXT and UP or DOWN to select and change the time and date. Then enter the new setting by
22 Functional verification Welch Allyn Propaq CS Vital Signs Monitor• Audible vital signs alarms (software versions 3.6x and greater): OFF (SETUP >
Service Manual Functional verification 2315. Press HOME > SETUP > ALARMS > STAT SET to automatically set heart and respiration rate alarm li
24 Functional verification Welch Allyn Propaq CS Vital Signs Monitor8. Move the wire and scope probe from the ECG pin of the ECG X1000 connector to th
Service Manual Functional verification 251. P r e s s SETUP > WAVE SELECT and check that the P1 and P2 waveforms are turned on. 2. Connect the IB
iiiContents1 - Safety summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Warnings. . . . . . . . . . . . . .
26 Functional verification Welch Allyn Propaq CS Vital Signs Monitornibp test screen7. Manually reduce the pressure as indicated on the pressure meter
Service Manual Functional verification 27Temperature1. Plug a 37.0° C YSI 400 series (two conductor) calibrated temperature probe simulator into the T
28 Functional verification Welch Allyn Propaq CS Vital Signs Monitor• Masimo-compatible SpO2 cable.• Compatible adult SpO2 sensor for finger.• BIO-TEK
Service Manual Functional verification 29“Expansion module option configurations” on page 8. The following table shows how to identify compatible sens
30 Functional verification Welch Allyn Propaq CS Vital Signs Monitor18. Close the paper door and set the monitor upright.19. Simultaneously press the
Service Manual Functional verification 31b. The reported ATIS is 30.09 inches Hg. (Altimeter: three zero zero niner.)c. Multiply the LUP (632.35) by t
32 Functional verification Welch Allyn Propaq CS Vital Signs Monitor-500 773.83 2500 693.80 7000 586.44 15000 428.89-400 771.05 2600 691.26
Service Manual Functional verification 33100 751.03 1800 611.22 3500 493.27 5200 394.46150 746.58 1850 607.45 3550 490.10 5250 391.82200
34 Functional verification Welch Allyn Propaq CS Vital Signs MonitorMainstream CO2 verification procedureThis procedure verifies the accuracy of the m
Service Manual Functional verification 35WARNING Eliminate all the possibilities causing errors shown in the following table before performing the c
iv Contents Welch Allyn Propaq CS Vital Signs MonitorPropaq CS Monitor software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
36 Functional verification Welch Allyn Propaq CS Vital Signs MonitorSidestream CO2 verification procedureThe following procedures verify the operation
Service Manual Functional verification 3711. Plug the water trap inlet and verify that the monitor generates an OCCLUSION-CHECK WATER TRAP/TUBING equi
38 Functional verification Welch Allyn Propaq CS Vital Signs Monitor30. Multiply the calculated partial pressure (in mmHg) from step 29 by 0.95316. Re
Service Manual Functional verification 39Functional verification for Acuity Wireless optionTo perform this procedure, you will need access to a proper
40 Functional verification Welch Allyn Propaq CS Vital Signs MonitorReset Radio menu (5 GHz radio card)For monitors containing a 5 GHz radio card, a c
Service Manual Functional verification 41Once the sequence is entered correctly, the Reset Radio menu will be displayed. Press YES to reset the radio
42 Functional verification Welch Allyn Propaq CS Vital Signs Monitor
443CalibrationIntroductionThis section provides procedures to perform all adjustments required to calibrate the Propaq CS monitor to conform to Welch
44 Calibration Welch Allyn Propaq CS Vital Signs MonitorEquipment neededYou will need the following equipment to calibrate the Propaq CS monitor. Part
Service Manual Calibration 455. Open the monitor. 6. Disconnect P1 (expansion connector cable) from the Analog Board and open the front chassis approx
11Safety summaryWarningsThis Safety Summary should be read by all Propaq CS monitor users. Specific warnings and cautions are placed throughout the do
46 Calibration Welch Allyn Propaq CS Vital Signs MonitorRecharger supply adjustments1. Verify that the monitor power switch is off. 2. Disconnect the
Service Manual Calibration 478. Verify that the input current from the variable dc power supply is less than 1.8 A and that the green LED on the monit
48 Calibration Welch Allyn Propaq CS Vital Signs MonitorMain power supply adjustmentsThe following procedures assume you are continuing from the Recha
Service Manual Calibration 498. Disconnect the DMM leads.Table 12. Main power supply voltage checksSupply voltages Ref. lead Pos. lead Voltage limits
50 Calibration Welch Allyn Propaq CS Vital Signs MonitorCalibrating realtime ECG out1. Connect the 5-lead ECG cable to the monitor and the simulator.2
Service Manual Calibration 51Noninvasive Blood Pressure calibration1. Connect the cuff, digital pressure meter, bulb, and the Encore together with the
52 Calibration Welch Allyn Propaq CS Vital Signs Monitor11. Connect the oscilloscope reference lead to TP43 or TP1 on the Analog Board.12. Check for ≤
Service Manual Calibration 53Invasive Blood PressureSkip this part of the calibration procedure if your monitor does not support IBP measurements. 1.
54 Calibration Welch Allyn Propaq CS Vital Signs MonitorMSP Board (MCO2) calibrationThe monitor MSP Board has two calibration adjustments. One adjustm
555TroubleshootingIntroductionThis section provides information that can be helpful in troubleshooting the monitor.Screen messagesMessages may appear
2 Safety summary Welch Allyn Propaq CS Vital Signs Monitor• To ensure conformance to risk (leakage) current requirements when operating from an ac mai
56 Troubleshooting Welch Allyn Propaq CS Vital Signs MonitorLACK OF WAVE, CHECK ADAPTER, SENSOR - (MCO2). The airway adapter (cuvette) may be contamin
Service Manual Troubleshooting 57Battery capacity checkSeveral variables affect monitor operating time on battery:• active options• frequency of NIBP
58 Troubleshooting Welch Allyn Propaq CS Vital Signs Monitor
659Repair proceduresIntroductionThis section provides instructions for removing and replacing serviceable modules in the:• Propaq CS monitor• Expansio
60 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorRequired toolsThe following tools are required to disassemble and reassemble the Propaq
Service Manual Repair procedures 61Propaq CS Monitor optionsYour monitor may contain an option module that attaches to the monitor. You must remove th
62 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorField-Replaceable Units in the monitorFront chassisThe Front Chassis of the monitor cont
Service Manual Repair procedures 63Interface chassisThe Interface Chassis is available either with or without the Nurse Call option.Rear chassisThe Re
64 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorReplacing the power input fuseThe Fuse F1 protects the Recharger Board circuitry against
Service Manual Repair procedures 65Keep the battery away from children. Do not disassemble it and do not incinerate or burn it. The battery maintains
Service Manual Safety summary 3• While under warranty, the monitor should be serviced only by a Welch Allyn service technician. DefinitionsSymbolsThe
66 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorRemoving the single battery packThe following steps describe how to remove the battery p
Service Manual Repair procedures 672. If necessary, use the flat blade screwdriver to gently loosen the Expansion Module from the monitor (see below).
68 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor7. Remove the dual battery pack from the Expansion Module.8. Store the battery in a safe
Service Manual Repair procedures 693. Slightly spread the casing open at the top until you can disconnect the pressure transducer tube indicated below
70 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorP1 removal6. Spread the casing open to an angle of about 45-60 degrees.7. Detach the fol
Service Manual Repair procedures 718. Detach the ribbon cable from the Rear Chassis (see below):Ribbon cable to P10 on the Recharger Board, from the D
72 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorClosing the monitor1. Before closing the monitor casing, confirm the following:• Cable t
Service Manual Repair procedures 73Replacing the Analog BoardThis section provides instructions on removing the Analog Board. The Analog Board is secu
74 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor6. Gently lift the Analog Board up from the Front Chassis.When installing the Analog Boa
Service Manual Repair procedures 75Replacing the DisplayThe Digital Board is located beneath the Analog Board. It is necessary first to pull off the b
4 Safety summary Welch Allyn Propaq CS Vital Signs MonitorKeep away from rain Altitude limitPatient connections are Type B The device has met all esse
76 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor7. Remove the nine screws connecting the Digital Board to the Display (see below).IMPORT
Service Manual Repair procedures 77Reinstalling the DisplayWhen installing the Display:• Make sure the cable between the Digital Board and Recharger B
78 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorReplacing Recharger Board fuse (F2)The Recharger Board fuse (F2) protects the monitor po
Service Manual Repair procedures 79Replacing air tubingReplace the NIBP tubing as a single assembly to ensure the tubing lengths are accurate. Refer t
80 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorReplacing cablesIf you suspect a cable is faulty, check its continuity on all pins. If y
Service Manual Repair procedures 811097-02BeveledCornerPROMSlotBeveledOutsideCornerSlotFingers
82 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorCuff fittingsAir system integrity in the NIBP system is essential to ensure accurate mea
Service Manual Repair procedures 83Replacing the Recharger BoardThe Recharger Board is secured to the Rear Chassis by four screws. When removing the R
84 Repair procedures Welch Allyn Propaq CS Vital Signs Monitor6. Lift out the Recharger Board.When installing the Recharger Board:• blow out any dust
Service Manual Repair procedures 85Replacing the pumpNew wires and a connector are provided with the replacement pump kit.1. Remove the battery pack (
25OverviewPurpose and scopeThe Propaq CS Service Manual is intended as a reference for monitor maintenance and repair to the field-replaceable unit (F
86 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorReplacing the side panelsThe Propaq CS monitor side panels are complete assemblies.The s
Service Manual Repair procedures 87Removing the left side panel1. Remove the battery pack (“Removing the single battery pack” on page 66).2. Open the
88 Repair procedures Welch Allyn Propaq CS Vital Signs Monitorside panel rou ting4. On the new side panel, remove the protective paper from the gasket
Service Manual Repair procedures 890696-45
90 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorExpansion ModuleThe Expansion Module housing contains one or all of the following (see b
Service Manual Repair procedures 91Opening Expansion Module rear panel1. Separate the monitor and Expansion Module (“Removing the dual battery in expa
92 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorClosing Expansion Module1. Make sure all cables are in place and the gaskets removed ear
Service Manual Repair procedures 93Replacing Expansion Module front panelThe front panel contains the printer buttons. Like the rear panel, the front
94 Repair procedures Welch Allyn Propaq CS Vital Signs MonitorReplacing Expansion Module button board and buttonsThe Expansion Module contains a butto
Service Manual Repair procedures 95Replacing the printerThis section provides instructions on removing and installing the printer mechanism and Printe