Technical Manual
1-4 Welch-Allyn, Inc.
Probe Warmer Circuit Self Tests
Proper instrument functionality should be verified first as described in the
Instrument
Reset/Self Tests
section (above) before a probe is installed. With a properly
functioning instrument, the probe can be run through the self test. If the instrument is
functioning properly and a probe is installed, the instrument will initiate the probe
warmer self test during the instrument reset self test and whenever a probe is
plugged in.
If an instrument has passed the
Instrument Reset/Self Tests
section, install the
probe connector and observe the display of the probe type. For Models 678 and 679,
the probe type will display for approximately 2 seconds.
If the display does not show “OrL” or “Aly” with an Oral/Axillary probe plugged in,
or “rEC” with a Rectal probe plugged in, there is a problem with either the probe or
the probe connector in the instrument.
If the display goes blank after the probe type display, the probe has passed its
tests and the instrument is ready for use. Do not withdraw the probe during this
self test.
If the display shows the malfunction icon |
X
|, refer to the
Troubleshooting
section.
Note:
Handle the probe only by the probe handle, not the metal shaft. When
removing a probe, disconnect the connector by pressing the locking tab and
pulling on the connector body. Do not pull it by the cord.
If there are any problems with probe initialization, refer to the
Troubleshooting
section.
Battery Life
Under normal use, battery life is expected to provide approximately 5,000 temperature
measurements. This number is based on a 22.2
°
C (72.0
°
F) ambient temperature,
with the security features turned off. Colder ambient temperatures, excessive security
alarms, use of non-alkaline batteries, and other usage patterns can reduce battery life.
Instruments are shipped with fresh batteries, but we cannot guarantee full life from the
first set of batteries due to potential long storage times between shipping and actual
use. Battery life can also be reduced by storage at elevated temperatures.
If you are experiencing short battery life, refer to
Battery Life Problems
in the
Troubleshooting
section.