Power Adapter Specifications
Characteristic Specification
Protection classifications, all adapters
a
b
a. Per EN 60601-1 unless otherwise stated.
b. See Welch Allyn Products and Accessories for model numbers.
Type of Protection Against Electric Shock Class I, (Protectively Earthed)
Degree of Protection Against Harmful Ingress
of Water
For ordinary, indoor locations only.
Method of Disinfection Not suitable for autoclaving
Flammable Anesthetics Not suitable for use with flammable anesthetics
Environmental specifications, all adapters
Operating Temperature 0° to 50° C
Shipping and Storage Temperature -20° to 60° C
Operating Altitude -2,000 to 15,000 feet (-610 to 4,572 m)
Shipping and Storage Altitude -2,000 to 40,000 feet (-610 to 12,192 m)
Operating Relative Humidity 15% to 95%, noncondensing
Shipping, Storage Relative Humidity 15% to 95%, noncondensing
Shock 50 g
Vibration Random Vibration, 0.02 g
2
/Hz from 10 to 300 Hz, ramping down to 0.002
g
2
/Hz at 500 Hz. Operating 1 hour per axis, 3 hours/ test.
Physical specifications
Length 5.0 in (12.7 cm)
Width 3.6 in (9.1 cm)
Height 3.1 in (7.9 cm)
Weight 3.1 lb (1.4 kg)
Universal power adapter, part No. 503-0054-00
Rated Input 100-120 V ac, 500 mA, 50/60 Hz
Rated Fuses T800 mA/250 V, Time-Delay, 5x20mm
Rated Output (Continuous) 16-24 V dc, 25 VA
Additional Features Detachable power cord, pilot light
Universal power adapter, part No. 503-0054-01
Rated Input 200-240 V ac, 250 mA, 50/60 Hz
Rated Fuses T400 mA/250 V, Time-Delay, 5 x 20mm
Rated Output (Continuous) 16-24 V dc, 25 VA
Additional Features Detachable power cord, pilot light