46 Monitoring Welch Allyn Propaq CS Vital Signs Monitor
Sidestream CO
2
accuracy decreases if additional tubing is connected to the sample
line. Avoid connecting additional tubing to the standard sample line.
Set up the CO
2
display and alarm limits
CO
2
monitoring is typically displayed as shown:
1. To adjust the display, from the Main Menu press SpO
2
/CO
2
, CO
2
to access the first
CO
2
menu:
WARNING The exhaust port for sidestream CO
2
is an output for the expired
gases from the patient and any connected breathing apparatus. The exhaust port
is intended only for connection to gas collection equipment such as gas
scavenger devices (the device should comply with ISO 8835-3:1997 E). Do not
allow any other connection to the exhaust port.
WARNING If using a watertrap, the watertrap is disposable and should only be
used for a single patient. Do not reuse the watertrap for another patient.
Note
After you connect a mainstream CO
2
sensor or sidestream CO
2
watertrap, the
Propaq CS monitor displays the waveform briefly without a scale. It displays
WARM UP (for mainstream) or START UP (for sidestream) in the CO
2
numerics
window. After about 30 seconds, the monitor displays the CO
2
measurement and
waveform range.
NET
OFF
HR
MON
17:05:10
PRINT FAULT
II
BR
P
1
( 96)
BPM
mmHg
(M)
TEMP
100.4
98.6
1.8
97
121
79
S
D
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
12
S
D
NIBP
ECG/
RESP
INVASIVE
PRESSURE
SpO
2
/
CO
2
SETUP
ADULT
ON NETWORK
T1
T2
ΔT
mV
cm
1
60
30
0
M
C
O
2
80
ALLEN KIM
MCO
2
mmHg
35
10
MCO
2
indicates mainstream CO
2
is active.
SCO
2
is displayed when sidestream CO
2
is active.
CO
2
is displayed if a mainstream sensor or sidestream watertrap is installed but not active.
ETCO
2
INCO
2
appears here when
displayed. The INCO
2
numeric value is only
displayed if it is in alarm or
if it is 7.5 mmHg (or 1kPa
or 1%).
Breath Rate is determined
from the CO
2
sensor.
Measured CO
2
levels are normally displayed as both a waveform and an ETCO
2
numeric value.
RANGE
mm/s
PREVIOUS
MENU
MORE