Directions for Use Acuity 77
Wireless connection to Acuity
Connect a new patient
1. Make sure the Acuity network cable is not plugged into the Acuity connector and an
Acuity bedside network jack. Turn on power to the Wireless Propaq CS. The monitor
displays the powerup screen for about 10 seconds, then displays the Main Menu. The
monitor is in the powerup patient mode with the associated settings.
2. Confirm that within a few seconds the Wireless Propaq CS displays the flashing
message WIRELESS in the upper left corner of the display, alternating with one of
these messages:
SEARCHING indicates the monitor is searching for a connection with an access point.
CONNECTING indicates the monitor has associated with an access point, but is not
fully connected to the Acuity network.
3. Confirm that within one minute the monitor displays these alternating messages:
ON NETWORK and WIRELESS
This indicates the Wireless Propaq CS is connected to the Acuity network.
4. After the network connection is established, the monitor may prompt you to select an
Acuity Unit (if your facility has more than one Acuity unit). Scroll up or down to
highlight the desired Acuity Unit, then press Select.
5. The Wireless Propaq CS displays a list of possible patients. If your patient has been
pre-admitted to the selected Acuity unit, they will be included in the list.
If your patient is NOT in the list, highlight Select Patient at Central and press Select.
The patient name will need to be entered later at the Acuity Central Station.
If your patient is in the list, scroll to highlight your patient’s name, then press Select.
6. Within a few seconds the Wireless Propaq CS displays a list of unassigned rooms.
If you want to assign the patient to a room, highlight the room and press Select.
If you do not want to assign a room at this time, highlight Select Room at
Central and press Select.
The patient room will need to be entered later at the Wireless Propaq CS (see
page 79) or at Acuity (see Acuity Directions For Use).
7. Customize alarm limits for your patient if needed.
WARNING If you do not select the patient name at the Wireless Propaq CS at
this time, do not adjust any alarm limits until after the patient name and ID are
confirmed at Acuity. When the patient name and ID are confirmed at Acuity,
Acuity downloads the default settings and patient alarm limits for that Acuity unit
to the Wireless Propaq CS, thereby overriding any previous settings and alarm
limits.