60 Alarms & Alerts Welch Allyn Propaq CS Vital Signs Monitor
waveform is displayed without range values. Replace the adapter or replace the
sensor.
LACK OF WAVE - CHECK ADAPTER, SENSOR. Either the airway adapter is
obstructed or the CO
2
sensor has failed. Replace the airway adapter if it is obstructed.
The sensor must be unplugged and plugged in again.
LOW BATTERY - HEATER DISABLED (UNCAL appears in the numerics area). The
monitor’s battery voltage is too low. The CO
2
waveform is displayed without range
values. To continue operation, supply ac power to the monitor.
NO MAINSTREAM SENSOR DETECTED (SRCH appears in the numerics area). The
mainstream CO
2
sensor has been disconnected from the Propaq CS monitor after
providing CO
2
values. Disconnect and reconnect the sensor to the monitor if
necessary.
NON-PROTOCOL SENSOR (UNCAL appears in the numerics area). A CO
2
sensor has
been connected that does not match Welch Allyns specifications. The CO
2
waveform
is displayed without range values. Replace the sensor with a Welch Allyn CO
2
sensor.
SENSOR FAILURE - CALIBRATION ERROR. A sensor is defective or out of calibration
and disabled. Replace the sensor.
SENSOR FAILURE - EEPROM. The sensor has failed. Replace the sensor.
SENSOR FAILURE - HEATER. The sensor’s temperature control circuit or the
monitor’s CO2 circuitry has failed. Try replacing the sensor. If the message reappears,
have the monitor serviced.
SENSOR FAILURE - MOTOR DRIVE. The sensor’s motor drive (in the sensor head)
has failed. Replace the sensor.
SENSOR TEMPERATURE TOO HIGH. The sensors temperature is too high. The
sensor’s ambient operating range is 10° to 46° C. When the ambient temperature
returns to this range, this message is automatically removed and operation is
restored.
The following messages can appear in the numerics display area.
OFF. No CO2 source is selected.
SRCH. The sensor is preparing for a measurement.
UNCAL. The monitor has detected a problem such as a lack of calibration, an
obstruction, or a low battery.
WARM UP. The sensor heater is warming up. Wait 20 to 30 seconds for the sensor to
heat. Values should appear in the numerics area when the sensor is sufficiently warm.
Sidestream CO
2
messages
ALTIMETER FAILURE - RANGE. The Propaq CS monitor is operating at an altitude
outside the sidestream CO2 options operating altitude range of -2,000 to 15,000 feet.
Returning the monitor to within this range automatically cancels this message and
restores operation.
ALTIMETER FAILURE - RATE. The altimeter has detected that the ambient pressure is
changing at a rate greater than 100 mmHg/minute. When the rate of change is back
within the 100 mmHg/minute range, disconnect and reconnect the CO2 sensor to the
monitor.