Commands and Responses SureTemp Plus Module OEM Implementation
38 Confidential Welch Allyn OEM Technologies
Commands and Responses
(2900) Send Serial Number
Causes the module to send its serial number to the host.
Command Code (hex)
2900
Data
Size (bytes) Type Description
0 N/A None
Response Data
Size (bytes) Type Description
11 ansichar Module Serial Number
(10 ASCII characters + null terminator)
(2913) Send Temperature Method
Causes the module to send to the host the current temperature method. This flag is used to
specify whether the current temperature-taking method is normal predictive or monitor.
Command Code (hex)
2913
Data
Size (bytes) Type Description
0 N/A None
Response Data
Size (bytes) Type Description
1 int8 0: Normal (predictive)
1: Monitor