Welch-allyn SureTemp PlusModule - User Manual Bedienungsanleitung Seite 44

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 85
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 43
Commands and Responses SureTemp Plus Module OEM Implementation
44 Confidential Welch Allyn OEM Technologies
(2A37) Set Default Thermometer Probe Mode
Causes the module to set the default probe mode from the host. This flag is used to define
what algorithm the device defaults to when an oral/axillary probe is connected to the device.
Command Code (hex)
2A37
Data
Size (bytes) Type Description
1 int8 0: Oral/Axillary
1: Pediatric Axillary
2: Adult Axillary
3: Use Last Algorithm
Response Data
Size (bytes) Type Description
0 N/A None
Notes
Use the “Save Calibration Data to EEPROM” command defined in for (2A06) Save Data to
EEPROM on page 53 to save permanently to EEPROM.
(3000) Restart Module
Causes the module to reboot.
Command Code (hex)
3000
Data
Size (bytes) Type Description
0 N/A None
Response Data
Size (bytes) Type Description
0 N/A None
Seitenansicht 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 84 85

Kommentare zu diesen Handbüchern

Keine Kommentare